BPI® Photometers
are sold with a calibration lens that is traceable back to the National
Institute of Standards and Technology. (NIST). Users should regularly
check the performance of their meters with these lenses to ensure
continuing accuracy. BPI® Photometers can be expected to make an accurate assessment of UV transmission because of the characteristics of UV lenses and UV absorbers. When treated with a UV absorber the natural cut-off point is moved into the region of the UVA that the photometer reads best. In an untreated CR-39™ lens, the error of a photometer could be 10% of the UV transmission reading. (A range of 3 to 4%). This may seem like a large range of error, however
it is insignificant because of the amount of UV that is being passed by
the lens. The UV transmission is clearly in the
‘Danger’ zone and the actual percentage of
transmission is immaterial. * DANGER: More than 13% transmission. CAUTION: 4 -
12% transmission. SAFE: Less than 3% transmission. |
|
|
|
|
BPI®
UV Safety |
BPI®
UV Economy |
BPI®
UV |
|
| BPI# 109513 (110v) | BPI# 109514 (110v) | BPI# 119514 (110v) | |
| BPI# 209513 (220v) | BPI# 209514 (220v) | BPI# 219514 (220v) | |
| Test lenses in or out of frames. | Test lenses in or out of frames. | Test lenses in or out of frames. | |
| Lens clearance: 0.75in | 2cm | Lens clearance: 0.75in | 2cm | Lens clearance: 0.75in | 2cm | |
| Weight: 6lbs | 2.72kg | Weight: 6lbs | 2.72kg | Weight: 6lbs | 2.72kg | |
| Height: 5in | 12.7cm | Height: 5in | 12.7cm | Height: 5in | 12.7cm | |
| Left to right: 6.25in | 15cm | Left to right: 6.25in | 15cm | Left to right: 4.75in | 12cm | |
| Front to back: 6.75in | 17.2cm | Front to back: 6.75in | 17.2cm | Front to back: 6.5in | 12.7cm | |
| Test range: 350 - 400nm | Test range: 350 - 400nm | Test range: 350 - 400nm | |
| Amperage: 1 amp | Amperage: 1 amp | Amperage: 1 amp | |
| Fuse: 1 amp/250v. | Fuse: 1 amp/250v. | N/A (Power pack) | |
| Calibrate
manually to an Indicator light |
Calibrate
manually until zero shows on the display |
Calibrate
manually until zero shows on the display |
|
| Display:
3 Indicator lights - DANGER CAUTION SAFE* |
Display:
Shows % of UV passing through the lens |
Display:
Shows % of UV passing through the lens |
|
|
Set-up kit: Calibration lens. Manual. Patient brochures and stand. |
|||
|
UL Certification. Meets/exceeds CE standards |
|||
|
|
|
|
|
|
BPI®
|
BPI®
Dual |
BPI®
Transchromatic |
|
| BPI# 109515 (110v) | BPI# 119511 (110v) ** | BPI# 119513 (110v) | |
| BPI# 209512 (220v) | BPI# 219516 (220v) ** | BPI# 219513 (220v) | |
| Test lenses in or out of frames. | Test lenses in or out of frames. | Test lenses in or out of frames. | |
| Lens clearance: 0.75in | 2cm | Lens clearance: 0.75in | 2cm | Lens clearance: 0.75in | 2cm | |
| Weight: 6lbs | 2.72kg | Weight: 6lbs | 2.72kg | Weight: 6lbs | 2.72kg | |
| Height: 5in | 12.7cm | Height: 5in | 12.7cm | Height: 5in | 12.7cm | |
| Left to right: 6.25in | 15cm | Left to right: 6.25in | 15cm | Left to right: 6.25in | 15cm | |
| Front to back: 6.75in | 17.2cm | Front to back: 6.75in | 17.2cm | Front to back: 6.75in | 17.2cm | |
| Test range: 350 - 400nm | Test range: 350 - 700nm | Test range: 350 - 700nm | |
| Amperage: 1 amp | Amperage: 1 amp | Amperage: 1 amp | |
| Fuse: 1 amp/250v. | Fuse: 1 amp/250v. | Fuse: 1 amp/250v. | |
| Automatic calibration | Automatic calibration | Automatic calibration | |
| Display: Shows % of UV passing through the lens. Also has 3 indicator lights: DANGER CAUTION SAFE* |
Display will alternate between showing % of UV and % of VISIBLE light passing through the lens. Also has 3 indicator lights: DANGER CAUTION SAFE* |
Display will alternate between showing % of UV and % of VISIBLE light passing through the lens. In photochromic mode, it will darken lenses and read UV and VISIBLE progressively. |
|
|
Set-up kit: Calibration lens. Manual. Patient brochures and stand. |
|||
|
UL listed components. Meets/exceeds CE standards |
|||
|
This analyzer has been designed to determine if tinted lenses meet EC standards for traffic signal safety as outlined in ISO 14889, ISO DIS 8980-3 and ISO 13666. The BPI® Traffic Signal Safety Analyzer™ is an invaluable aid for quality control of lenses tinted with visible light treatments. The meter may be used for fashion tint control and for critical density adjustments on therapeutic tints such as BPI® Diamond Dye™ 500/550 and BPI® Diamond Dye™ 540. It can be used to check for UVA blocking. It may also be used for traffic signal color density and general filter level checks. It has a serial connection for data storage and printing. FILTER LEVEL LED’S 5 Filter Level LED's are located on the left
side. |
|
|
BPI®
Traffic |
|
| BPI# 119512 (110v) | |
| BPI# 219511 (220v) | |
| Test lenses in or out of frames. | |
| Lens clearance: 0.75in | 2cm | |
| Weight: 6lbs | 2.72kg | |
| Height: 5in | 12.7cm | |
| Left to right: 6.25in | 15cm | |
| Front to back: 6.75in | 17.2cm | |
| Test range: 350 - 700nm | |
| Amperage: 1 amp | |
| Fuse: 1 amp/250v. | |
| Automatic calibration | |
| Display: Red: % transmission at 610nm Yellow: % transmission at 580nm Green: % transmission at 558nm Blue: % transmission at 470nm White: % transmission of visible light 400-700nm UV: % transmission of UV light 350-400nm |
|
| Set-up kit: Calibration lens. Manual. Patient brochures and stand. | |
| UL listed components. Meets/exceeds CE standards |
|
The BPI UV Activator™
is a dual-purpose device which can be used as a demonstration of the
effectiveness of UV blocking lenses and also to activate photochromic
lenses with ultraviolet light of intensity similar to that in
sunlight. The UV light source is discreetly baffled inside the black
metal box away from the direct view of the user. It illuminates a white
target which fluoresces brightly in the presence of UV light. A
UV-blocking lens will leave a dark shadow on the fluorescent target.
Photochromic lenses will be darkened by the UV light at much the
same rate as they would be by natural sunlight. Lenses may be evaluated
while still in the frames.
|